Products
OPS Series Linear Programmable DC Power Supply
OPS Series is high efficiency, high performance programmable DC power supply with RS232C&GPIB(IEEE-488.2, option) interface based on SCPI(Standard Commands for Programmable instrument) protocol.
In addition, It is designed to be equipped in 3U x Half of 19inch wide rack.
Dimension
90W~20kW(1CH) : 213mm(W) x 370mm(D) x 132~1800mm(H)Others : 19’ Rack Cabinet Size (Please contact us to inquire about it)
-
1 Over Voltage Protection Key 14 Error message Display Key 2 Over Current Protection Key 15 Menu Excape Key 3 I/O Config or LOCAL Key 16 Voltage/Current Cursor or Menu Key 4 LOCK Key 17 Voltage/Current Cursor or Menu Key 5 Store present condition or Calibration Key 18 Main Power ON/OFF Switch 6 Restore memory save slot or Factory Key 19 Remote V+ Sensing Input Terminal 7 Output Voltage/Current ON/OFF 20 + Output terminal 8 Choice Voltage/Current or Limit Display Key 21 – Output terminal 9 Voltage/Current/Numbering Encorder Key 22 Remote V- Sensing Input terminal 10 Cycling mode˙s STEP Setting Key 23 Earth GND terminal 11 Cycling mode˙s Sequencne Control Key 12 Cycling mode˙s Repeating Key 13 Cycling mode˙s RUN/STOP Key - 5-Digit Display Resolution
- 16×1 Big-Character Type VFD Display
- Adjustment Encoder For Voltage and Current setting
- CV/CC Mode
- Remote Voltage-Sensing(Voltage drop compensation)
- OVP, OCP Protection Mode
- Output ON/OFF Function
- Panel Lock Function
- Memory Save and Recall Up to 10 operation states
- Save and Check up to 10 error messages
- Self-Diagnosis Test Mode
- Cycling Mode(Voltage, Current slope and Delay time up to 100 slots)
- Calibration Function(Manual and PC)
- Factory Mode(Last state, Calibration back-up etc)
- Half of 19inch wide x 3U Size(Under 300W)
- RS-232C Communication(GPIB is optional)
- Single Phase, AC Input 220VAC ±10%
-
-
Let˙s use the powerful store function!
-
OPS Series can save and recall up to 10 voltage and current, OVP, OCP that the user sets up. Even when a user is not an engineer, the user can use it for the production and product inspection, reliability test with the function for recalling the pre-set and stored data.
-
-
-
How does our company˙s productivity rank in the industry?
-
Our Programmable DC Power Supply is ideal for the FA system development since it can judge at least 30 equipments whether they are pass or fail within one second. It is possible to provide fine and precise source one using this product alone, since it has built-in 24 bit ADC. Likewise, it measures even the fine voltage and electric current. Thus, it is possible to reduce budget since it does not require separate voltage/ current measurement device at the outside.
-
-
-
-
STABILITY
-
Experience the upgraded stability
Even when the equipment˙s internal temperature changes or even when there are external factors that influence while conducting test for a long time due to the Temperature Compensation Circuit and real-time measurement Compensation Circuit that are applied to the OPS Series, it is possible to conduct test in an accurate and stable manner while maintaining the set up voltage.
-
-
-
PROTECTION
-
Now, use it with peace of mind. [Protection Mode]
Using the over-voltage (O.V.P) and over-current (O.C.P)mode that are built in the OPS Series blocks the output over-voltage or over-current to protect the source of over-load, and it can be used safely even in case of user negligence or when conducting test for a long time. Moreover, it is possible to reduce the defect rate of the production line by blocking malfunction caused by user˙s negligence.
-
-
-
Why? Why is the output voltage different from the set up voltage?
-
Remote Voltage Sensing
-
Power supplier that does not provide V-Sensing may not offer an accuracy level that a user wants. ODA Technologies˙s Programmable DC Power Supply provides V-Sensing (Remote Voltage Sensing) function to produce compensation output when it comes to the voltage that is damaged due to the cable˙s specific resistance property in order to supply increasingly precise power.
-
-
-
Now, measure immediately without delay after the output setting!
-
Competitor˙s products require delay for a specific period of time in order to obtain accurate measurement value after the setting order is communicated. Even when the order for setting is communicated, Analog Controller˙s responsiveness is late. Thus, error may result in which measurement takes place while output voltage is increasing. Meanwhile, it is necessary to collect data impeccably only when measurement takes place after waiting for the setting of the output voltage and electric current is completed. Our company˙s programmable power supply can work with 2 ms to conduct 15 V setting in the 0V.
-
-
-
The work can be carried out even when there is no program developer present.
-
Cycling Mode TEST!
-
Cycling Mode that is provided to the OPS Series provides memory so that up to 100 gradient and maintenance time of the voltage, electric current can be stored. When cycling Mode is used, it is possible to develop the system for the long-time reliability test for the product with ease without producing the Windows Application. Moreover, it is possible for the voltage and electric current to supply powers that are different in a regular, consistent manner to the mechanical device and facility besides the reliability test.
-
-
-
Pyramidal wave materialization -
It is possible to materialize the output of the pyramidal wave by receiving the Arbitrary Function Generator˙s input voltage with the Analog Input.
-
Line-Up and Specification
Current
Voltage1A
2A
3A
5A
10A
20A
30A
9V
–
–
–
–
OPS-910
OPS-920
OPS-930
18V
–
–
–
OPS-185
OPS-1810
OPS-1820
OPS-1830
30V
–
–
OPS-303
OPS-305
OPS-3010
OPS-3020
OPS-3030
50V
–
–
OPS-503
OPS-505
OPS-5010
OPS-5020
OPS-5030
80V
OPS-801
OPS-802
OPS-803
OPS-805
OPS-8010
OPS-8020
OPS-8030
100V
OPS-1001
OPS-1002
OPS-1003
OPS-1005
OPS-10010
OPS-10020
OPS-10030
150V
OPS-1501
OPS-1502
OPS-1503
OPS-1505
OPS-15010
OPS-15020
OPS-15030
200V
OPS-2001
OPS-2002
OPS-2003
OPS-2005
OPS-20010
OPS-20020
–
250V
OPS-2501
OPS-2502
OPS-2503
OPS-2505
OPS-25010
OPS-25020
–
300V
OPS-3001
OPS-3002
OPS-3003
OPS-3005
OPS-30010
–
–
Current
Voltage50A
80A
100A
150A
200A
9V
OPS-950
–
OPS-9100
OPS-9150
OPS-9200
18V
OPS-1850
OPS-1880
OPS-18100
OPS-18150
OPS-18200
30V
OPS-3050
OPS-3080
OPS-30100
–
–
50V
OPS-5050
OPS-5080
OPS-50100
–
–
80V
OPS-8050
–
–
–
–
100V
OPS-10050
–
–
–
–
150V
–
–
–
–
–
200V
–
–
–
–
–
250V
–
–
–
–
–
300V
–
–
–
–
–
Model
ParameterOPS-303
OPS-305
OPS-307
OPS-3010
Output rating 0~30V, 0~3A
0~30V, 0~5A
0~30V, 0~7A
0~30V, 0~10A
Resolution
Programming/Readback
≤ 250㎶ / ≤ 30㎂
≤ 250㎶ / ≤ 50㎂
≤ 250㎶ / ≤ 70㎂
≤ 250㎶ / ≤ 70㎂
Display Meter
1mV / 100㎂
1mV / 100㎂
1mV / 1mA
1mV / 1mA
Programming Accuracy (@25℃ ±5℃)±(%of output + offset)
Voltage
0.05% + 10mV
0.05% + 10mV
0.05% + 10mV
0.05% + 10mV
Current
0.15% + 5mA
0.15% + 5mA
0.2% + 10mA
0.2% + 10mA
Read-back Accuracy (@25℃ ±5℃)±(%of output + offset)
Voltage
0.05% + 5mV
0.05% + 5mV
0.05% + 5mV
0.05% + 5mV
Current
0.08% + 3mA
0.08% + 3mA
0.15% + 5mA
0.15% + 5mA
Ripple & Noise
≤ 2mVp-p, ≤ 2mArms
≤ 2mVp-p, ≤ 2mArms
≤ 2mVp-p, ≤ 2mArms
≤ 2mVp-p, ≤ 2mArms
Load Regulation
2mV, 500㎂
2mV, 500㎂
2mV, 500㎂
2mV, 500㎂
Line Regulation
500㎶, 500㎂
500㎶, 500㎂
500㎶, 500㎂
500㎶, 1mA
Command Processing Time
<32ms
<32ms
<32ms
<32ms
Voltage Programming Speed(No load)
Rising time
≤ 7.5V/㎳
≤ 7.5V/㎳
≤ 7.5V/㎳
≤ 7.5V/㎳
Falling time
≤ 3V/㎳
≤ 3V/㎳
≤ 3V/㎳
≤ 3V/㎳
Dimension(WxHxD/mm)
213x132x370
213x132x370
213x132x370
213x132x370
Transient Response time
Less than 50㎲ for output to recover to within 15mV following a change in output current from full load to half load or vice versa
Model
ParameterOPS-503
OPS-505
OPS-801
OPS-802
Output rating 0~50V, 0~3A
0~50V, 0~5A
0~80V, 0~1A
0~80V, 0~2A
Resolution
Programming/Readback
≤ 500㎶ / ≤ 30㎂
≤ 500㎶ / ≤ 50㎂
≤ 800㎶ / ≤ 10㎂
≤ 800㎶ / ≤ 20㎂
Display Meter
1mV / 100㎂
1mV / 100㎂
1mV / 100㎂
1mV / 100㎂
Programming Accuracy (@25℃ ±5℃)±(%of output + offset)
Voltage
0.05% + 12mV
0.05% + 12mV
0.05% + 35mV
0.05% + 35mV
Current
0.15% + 5mA
0.15% + 5mA
0.15% + 5mA
0.15% + 5mA
Read-back Accuracy (@25℃ ±5℃)±(%of output + offset)
Voltage
0.05% + 12mV
0.05% + 12mV
0.05% + 35mV
0.05% + 35mV
Current
0.15% + 5mA
0.15% + 5mA
0.15% + 5mA
0.15% + 5mA
Ripple & Noise
≤ 3mVp-p, ≤ 2mArms
≤ 3mVp-p, ≤ 2mArms
≤ 6mVp-p, ≤ 2mArms
≤ 6mVp-p, ≤ 2mArms
Load Regulation
2mV, 500㎂
2mV, 500㎂
3mV, 500㎂
3mV, 500㎂
Line Regulation
500㎶, 500㎂
500㎶, 500㎂
1mV, 500㎂
1mV, 500㎂
Command Processing Time
<32ms
<32ms
<32ms
<32ms
Voltage Programming Speed(No load)
Rising time
≤ 7.5V/㎳
≤ 7.5V/㎳
≤ 7.5V/㎳
≤ 7.5V/㎳
Falling time
≤ 3V/㎳
≤ 3V/㎳
≤ 3V/㎳
≤ 3V/㎳
Dimension(WxHxD/mm)
213x132x370
213x132x370
213x132x370
213x132x370
Transient Response time
Less than 50㎲ for output to recover to within 15mV following a change in output current from full load to half load or vice versa
Common details
RS-232C Standard
Option
GPIB communication, GPIB cable, RS-232C cable, Output ON/OFF signal control, AC Power Cord (Other type), AC Input 100V ± 10%, 230V± 10%, 50~60Hz,
Rack Mount Support (Bracket and Shelf)-
Options
- GPIB communication
- GPIB cable
- RS-232C cable
- Output ON/OFF signal control
- AC Power Cord(Other type)
- AC Input 100V ± 10%, 230V± 10%, 50~60Hz
- Rack Mount Support(Bracket and Shelf)
-
Accessories
- AC Power Cord(Type F)
- Operating Manual(QR Code)
- Demo software program(QR Code)
- Banana to Alligator DC Output Cables(+,-)
-
-